Counter measurements (Hz)
Ranges
Range selection
Resolution
Number of representation units
Accuracy for counter measurements
Gate time
!.
Conversion rate
Trigger mode
Counter sensitivity
kHz
10
100
MHz
1
10
Range 100 kHz, 1 MHz and 10 MHz; Manual or automatic
Range 10 kHz; Manual only
0,1 Hz in range 10 kHz
99999
Range
Accuracy
10 kHz
100 kHz
1 MHz
± (0,005 % of reading +0,001 % of range)
10MHz
± (0,01 % of reading + 0,001 % of range)
Range
Gate time
10 kHz
10
s
100 kHz
1
s
1 MHz
100 ms
10 MHz
10 ms
Range
Rate
10 kHz
1 conv./IO s
100 kHz
1 MHz
10 MHz
1 conv./s
On positive going crossings of the + trigger level
•%
On negative going crossings of the — trigger level
DC to 1MHz
150 mV peak
1MHz to 10 MHZ
300 mV peak
Resolution;
1 mV
Trigger level adjustment
With thumbwheel at the front of the PM 2521