Chapter 1 Overview
1.3 Specifications
1.3
Specifications
Table 1.1
shows the specifications of the power synchronized unit and the AC sensor unit.
Table 1.1
Specifications
Dimensions (W D H)
Weight
Power
Maximum power requirements
Maximum heat generation
Connection interface
Number of
connectable
devices
Cable length
Conditions to detect AC output
Support for a duplicated configuration
Environmental condition
FUJITSU Storage ETERNUS DX, ETERNUS AF Configuration Guide (Power Synchronized Unit)
Item
Power synchronized unit
AC sensor unit
Power synchronized unit
AC sensor unit
Voltage
Phase
Frequency
AC100 - 120V
AC200 - 240V
AC100 - 120V
AC200 - 240V
For ETERNUS DX/AF storage systems
For AC detection
Basic
Server UPS unit
Maximum (when AC sensor units
are connected)
Basic
ETERNUS DX/AF
Maximum (when a cascade
connection is used)
Between a power synchronized unit
and an AC sensor unit (total)
Between a power synchronized unit
and the ETERNUS DX/AF
Between power synchronized units
Power consumption
Voltage
Phase
Temperature
Humidity
23
Copyright 2017 FUJITSU LIMITED
Specifications
440 230 40 (mm) [1U]
100 150 40 (mm) [1U]
4.0kg
1.3kg
AC100 - 120V, AC200 - 240V
Single
50Hz/60Hz
3W
7W
11kJ/h
26kJ/h
RS232C 4
AC sensor 2
2
16
4
8
Max. 100m
Max. 5m
Max. 10m
0.1A or less AC output for one second or
more is detected.
AC100V / AC200V
Single
Yes
5 - 40C (during operation)
0 - 50C (during non-operation)
20 - 80%RH (during operation)
8 - 80%RH (during non-operation)
P3AM-8012-10ENZ0