Siemens Maxum II Reference Manual page 64

Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

Electronic Compartment Component Descriptions and Maintenance Procedures
2.5 Detector Personality Modules
Detector Control Paths
Several control signals are available to control various detector functions as shown below.
Network
GCP
SYSCON
Color
Touch
Display
Figure 2-29
Maxum II Detector Control Functions
Location ID Switch
The Location ID Switch, shown previously in the photograph, selects the DPM location that is
incorporated in the address, to be reported in the results.
The DPM I
distribution board. In this scenario, the following values are applied:
NOTE:
If the DPM I
value is determined by the SNE.
64
2
2
I
C
I
C
BaseDPM
PECM
C port is connected directly to the system controller via the PECM or a wiring
2
Switch Value
1
2
3
C port is connected to an SNE, the value is always set to "1". The actual location
2
Digital
(Range
Select)
Mezzanine
Module
A/D
(signal conditioning)
Digital (Ignite) / Power (Glow Plug)
300 V Bias (FID)
Enable (FPD)
5 V Power (FPD)
FID
FPD
Analog
Output
Location
Left
Center
Right
Maxum II Reference Manual
Manual, 7/2017, 2000596-001