16 .
Change Signal Generator output for 50 MHz and amplitude level for 8 division vertical
display on HP 54501A .
17 .
Change HP 54501A TIMEBASE to 10 ns/div .
a.
Press front-panel TIMEBASE key .
b.
Select 10 ns/div with knob or ENTRY keys.
19 .
Triggering should be stable down to 1 division of signal amplitude .
20 .
Repeat steps 1 through 20 for Channel 4.
21 .
Repeat steps 12 through 20 for Channels 2 and 3.
HP 54501A PERFORMANCE TESTS
18.
Slowly decrease Signal Generator amplitude while maintaining stable triggering on HP
54501 A.