Transfer of photometry-assay calibration data (from analyzer to host)

Composition of text
STX
:
G
(1)
(1)
(2)
STD data 6
(32)
1)
Frame character (1 character)
':' is transferred.
2)
Function character (2 characters)
'G
' is transferred.
3)
Channel No. (3 characters)
2
1
10
10
4)
STD count (1 character)
STD count is any of '1' to '6' and variable according to the calibration method.
When STD count is '1,' STD data is followed immediately by SD value
information.
5)
STD data 1 to 6 (32 characters each)
The data for each STD has the composition below.
5
10
0
10
Each absorbance data is right-justified and preceded by one more spaces.
6)
Calibration alarm (1 character)
Refer to the data alarm code list.
Channel No.
STD count
(3)
Each parenthesized numeral indicates the byte count.
SD value information
(8)
Transfer sequence
0
10
Transfer sequence
0
10
1st absorbance data (6 bytes)
STD type (any of STD 1 to 6)(1 byte)
16 - 35
Calibration alarm
(1)
(1)
ETX
BCC
(1)
A test code in photometry-assay calibration is
indicated in a three-digit integer. Test numbers
are "bb1" to "b36" which correspond to test
codes in the analyzer.
Data alarm (1 byte)
2nd initial absorbance data (6 bytes)
2nd absorbance data (6 bytes)
1st initial absorbance data (6 bytes)
STD data 1
(32)
(Variable)
(1)
5
10
Prozone value (6 bytes)