Operation
13.2 General advice for the operation of the METPOINT
1. A lack of throughflow can lead to an increase in the residual oil content values.
2. When the system is unpressurised or when there is no compressed-air throughflow, this will be signalled. The
screen is red and displays a warning regarding the cause of malfunction, as well as measures for the elimination of
faults (see Chapter 12.3).
3. The power supply of the sensor unit must always be connected. Only in this manner can it be guaranteed that the
reference gas generation system functions reliably.
4. Read out the measured values of the device at regular intervals of the device's memory by means of the software.
Check these values with regard to temporary exceeding of the limit value.
5. The indication of the measured value is updated every two minutes. During this time, measurement continues in
the background.
6. Total reset of the device can only be effected through interruption of the power supply.
Please make sure that temperature variations during the operation of METPOINT® OCV are not too
strong (or even exceed the admissible temperature range).
The following operating conditions are monitored and indicated on the display as clear text:
Operating pressure too low
Measurement range exceeded
Calibration required
Sensor test required
You will find the respective indications regarding the aforementioned operating conditions in Chapter 12.3.