Sievers 900 and 5310 C On-Line TOC Analyzer Installation Instructions (continued)
Step 6 C
Install the Analog outputs and alarms
(continued).
Table 3: Serial and 4-20 mA
Inputs and Outputs (TB3)
Pin Number
Output
(from left)
12
4-20 mA ( - output)
11
4-20 mA ( + output)
10
Not used
9
Not used
8
Reserved
7
Reserved
6
Reserved
5
Not used
4
Serial (In)
3
Serial (Out)
2
Serial (Ground)
1
Not used
Step 7
Connect the inlet system and waste outlet.
1. Connect the 1/4" Teflon tubing with the capsule or inline
filter to the sample inlet on the iOS System. Tighten 1/4 turn
past finger-tight with a 9/16" open-end wrench. Do not over-
tighten the nut.*
* For raw water and heavy particulates applications, additional
pre-filtration is required. Please contact GE Analytical Instruments
if you need recommendations.
2. Connect the 3/4" OD waste line tubing to the waste outlet on
the iOS System by sliding the tubing over the barb fitting.
Where to Find Operation and Maintenance Manuals
For more detailed instructions, you can download operation and maintenance manuals for the Sievers 900 On-Line Analyzer, the Sievers
5310 C On-Line TOC Analyzer, and the 900 Inorganic Carbon Remover (ICR) from our Web site. Go to www.geinstruments.com, click Library,
then Manuals, and make your selection.
Americas: T 800 255 6964, 303 444 2009 | geai@ge.com
Europe/Middle East/Africa: T 44 (0) 161 864 6800 | geai.europe@ge.com
Asia Pacific: geai.asia@ge.com
www.geinstruments.com
©2016. General Electric Company. All Rights Reserved.
Step 8
Configure basic Analyzer settings.
Before using the Analyzer, you will need to configure various basic
settings. Some of these settings will not need to be changed again,
unless you move the Analyzer or reconfigure the operational environ-
ment. Refer to the 900 or 5310 C On-Line Operation and Maintenance
Manual for complete details.
Step 8 A
Power the Analyzer on.
Step 8 B
Enable password protection.
Step 8 C
Set the clock and time zone.
Step 8 D
Name the Analyzer location (optional).
Step 8 E
Set reagent expiration dates.
Step 8 F
Set up the data history.
Step 8 G
Set up the printer (pptional).
Step 8 H
Export or print constants.
Step 8 I
Set up data I/O.
Step 9
Flush the Analyzer
Before placing the Analyzer into normal operation, you should
perform a reagent flush to remove any gas bubbles that may
have formed in the reagent lines; you should then leave the
Analyzer powered on for 12 hours to allow the DI water reservoir
to properly complete rinse down action.
DQS 96003-01 Rev A
Sievers 900 and 5310 C
On-Line TOC Analyzers
Quick Start Guide
IMPORTANT
For more detailed instructions, you can download manuals for the Sievers 900 On-Line TOC Analyzer, the
Sievers 5310 C On-Line TOC Analyzer, or the Sievers Inorganic Carbon Remover (ICR) from our Web site. Go to
www.geinstruments.com, click Library, then Manuals, and make your selection.
System Specifications*
Linear range (900)
0.03 ppb - 50 ppm TOC
Linear range (5310 C)
4 ppb - 50 ppm TOC
Precision
<1% of RSD
Accuracy
±2% or ±0.5 ppb, whichever is greater
Analysis time
4 minutes
Sample flow rate
Analysis mode: 0.5 mL/min
(nominal)
Fast Flush (between samples): 1.0 mL/min
Required sample line
30 mL - 300 mL/min (for On-Line Mode)
flow rate
100-240 ± 10% VAC, 50 watts, 50/60 Hz
Power requirements
Fuses
No user-replaceable fuses; built-in circuit
breaker
Sample temperature
1 ºC to 95 ºC**
Ambient temperature
10 ºC to 40 ºC
Sample pressure
Up to 250 psig**
(in iOS System)
Normal operating
Intended for indoor use only; avoid
environment
sunlight
* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
** If the sample temperature and pressure are above 60 ºC and 100 psi, the 900 and 5310 C On-Line Analyzer models with the stainless steel iOS is required.
Maximum relative
Up to 95%, noncondensing
humidity
Maximum altitude
3,000 m (9,843 ft)
Calibration stability
Typically stable for 12 months
Chemical reagents
300 mL acid; 300 mL or 150 mL oxidizer
(prepackaged)
Outputs
Serial (RS-232), USB, parallel printer port,
4-20 mA output, two alarms, binary output
Installation/Overvoltage
II
category
Safety certificates
CE, ETL listed. Conforms to UL Std. 61010-1
Certified to CSA C22.2 No. 61010-1
Pollution degree
2
Display
Color, touch-sensitive LCD
Size
62.4 cm height x 45.2 cm width x 26.4 cm
depth (24.6 in x 17.8 in x 10.4 in)
Weight
16.9 kg (37.2 Ibs)
IP Rating
IP 45

Advertisement

loading

Summary of Contents for GE Sievers 900

  • Page 1 No user-replaceable fuses; built-in circuit breaker For more detailed instructions, you can download operation and maintenance manuals for the Sievers 900 On-Line Analyzer, the Sievers Pollution degree 5310 C On-Line TOC Analyzer, and the 900 Inorganic Carbon Remover (ICR) from our Web site. Go to www.geinstruments.com, click Library, Sample temperature 1 ºC to 95 ºC**...